26
position is then selected, it is possible that the
input signal may no longer match the display. In
this case (square-wave signals), linear interpol-
ation (ITPL LIN) is ideal for display purposes.
13.6 AMAG = ALT MAG mode
Keep pressing the MENU button until "AMAG
OFF" (or AMAG ON) appears at the top edge of
the screen in the middle. In this mode the
SELECT button can only be used to make a set-
ting if the Hold key has previously been pressed.
A M A G O F F
13.7 Go-NoGo mode (signal assessment)
Keep pressing the MENU button until "G-NG =>
XH OH" appears at the top of the screen in the
middle. With this function, the SELECT button
can be used to toggle between XH= Comparison
and OH = Hold.
G - N G => X H O H
(14) X-Y mode
The procedure to make the settings for the X-Y
function in storage mode is as follows:
a) Set the Vertical switch to the DUAL position
b) Press the HOLD button in order to pause sam-
pling. Both input signals (X = CH1 and Y =
CH2) continue to be displayed.
c) turn the Time/div switch to the right-hand
stop in the case of the OS-3020 D or press the
X-Y button in the "HORIZ DISP" panel on the
OS-3060 D.
Note:
If the brightness is too high, reduce the intens-
ity because otherwise this will damage the
sensitive internal coating of the CRT screen and
the signal could be permanently burnt into the
coating.
If one of these two functions (XH or OH) was sel-
ected, the basic value = 0.4 div must be set by
pressing the RECALL button once. If a higher
value is required, keep the SELECT button pres-
sed until the desired value is reached.
G - N G 0 . 4 d i v
0.4 div: the assessment area covers ±0.4 div
0.8 div: the assessment area covers ±0.8 div
1.2 div: the assessment area covers ±1.2 div
13.8 Header text line at top edge of screen with-
out MENU button
Assuming that all the necessary settings have
been made and the MENU button is no longer
required, keep this button pressed until the LED
under the button cap goes out. The header text
line at the top edge of the screen (right-hand
half) might look as follows:
If this field contains "SM",
smoothing is switched on,
if it is blank, SM is off
N O R M A V 1 6 S A S M
EQUIV in the range from 0.2 µs/div (OS-3060 D: 0.1 µs/div) to 2 µs/div
NORM in the range from 5 µs/div to 2 ms/div
ROLL in the range from 1 ms/div to 0.2 s/div or 20 s/div (in X100 roll mode)
If this line is blank, averaging is switched off;
in this case AV 16 = 16 passes for one display
SA = recording memory A
SB = recording memory B