(..0 Addition and difference measurements
The measurement
of
the addition and difference is
it
function
repres
entin
g one waveform by adding two signals. The operation of
the addition (ADD) represent s the algebraic sum of the CHI and CH2
signals, and the operation of the difference
represents the
algebraic difference between the CHI and CH2 signals.
Measuring procedure
of
ADD
of
this instrument,
OS
·
.1I1
(
~
)I)SRS
i
-,
as
follows :
I. Set up per paragraph 2-2-4(2) dual-trace meas urement.
2. Set both V
OLT
S
/DIY
switch [ 13J and [ 14J to the same
pOS
1l
10
n
and turn the VARIABLE controls [ 15J and [ 16J fully clockwise
until being click-stopped. Where the amplitude difference
of
the two signals is considerably large, reduce both VOLTS
/DIY
switches simultaneously as much as to make the amplitude of the
l
ar
ger signal be within the screen displ ay.
3. Select the
TRIG
GER
switch having the biggest signal.
4. Set the V.
MOD
E switch [ 19] to
ADD
position)
Then, the algebraic sum of the CH I and CH2 signals appea rs as a
signal waveform. At this time, since the position cha nge of
Vertical POSITION c
ont
rols [17] and [18J varies the measurement
values, operation shall be prohibited.
(Not
e)
If the two i
npu
t signals have the same phase, the two
signals ap
pear
as the algebraic
sum
of
the individual
traces (e.g. 4.2 DIY
+
1.2
DIY = 5.4 DIY).
Where the two
input
signals have 180
0
counter-phase,
the two signals
app
ear as the difference (e.g. 4.2 DIY
1.2 DIY
= 3.0 DIY).